Detector: | PID: Levels: 0.1 ppm to 0.01 ppm for hydrocarbons sensitive to PID, or FID: Levels: 0.5 ppm to 0.1 ppm for hydrocarbons sensitive to FID cycle time 1 to 10 min |
Range: | 0 - 1000 ppm for FID, 0- 100 ppm for PID |
Reproducibility: | <1% of FS |
GC specification: | Loop 0.075 to 2 ml, 10 port valve Synspec, analytical column 2 -13 m backflush column 0.5 to 2 m |
Dimensions: | 19” rack, 3 standard Height Units, depth 37.2 cm net, weight 12 Kgs |
Calibration: | Internal calibration switch for calibration with zero and span gas, gas stream required 25 ml at ambient pressure |
Consumption of gas: | PID / FID: nitrogen, quality 5.0, 2.0 bar, 5 ml/min FID: instrument air: dry and clean, 2.5 bar, 250 ml/min FID: hydrogen, quality 5.0, 2,5 bar at 20 ml/min |
Power demand: | 220 V AC, 200 VA (110 V AC available) |
Included hardware: | Computer Pentium class, harddisk >20 Gb, 6 “ full colour LCD display, I/O 2 x RS232, 4 x USB, Ethernet, PS2 keyboard/mouse, Touchscreen |
Approvals: | CE approval for EMC conformity: EN 61010-1, EN 61000-6-2 and EN 60111-6-3 |
Included software: | Windows98, GC software, TWA calculation modules, alarm setting modules |